Automatic time (min. 10 minutes /
max. 120 minutes)
Time interval from one calibration phase to the next one.
Measurement time (max. Automatic time minus 5 minutes)
Time interval within the automatic time in which the instrument gath-
er measurement values, i.e. in which the measurement gas will be
sucked. The difference between the measurement time and the au-
tomatic time is used for purging the sensors with fresh air.
Save to MMC (min. 1 second / max. 255 seconds)
Adjustment of the interval time for data logger recordings.
CSV+Header (<F1> for YES / <F4> for NO):
Adjustment of data logging with or without column headings.
Start the automatic measurement by setting the cursor to line "Au-
tomatic" and pressing the key <OK> (on top right on the display
appears "A" for automatic).

12.3. Data logger

Here a Data logger record ("Data logger" and key <OK> / on top
right on the display appears the disk symbol) can be started or fin-
ished (just available when using the multi-media card). With
<Memory> you can interrupt and continue the record (only in gas
analysis). For each recording one file will be written on the card. The
files will be numbered consecutively (J2KDL-00.csv, J2KDL-01.csv
and so on) and can be transferred to PC with a card reader. The
length of a dataset is 500 byte which means that on a 32 MB card
64000 measurements could be recorded.