5.5. Online configuration
General
DE Configure
Calibration
78
DE Main Menu Procedures
(ST 3000 / ST 800)
ST 3000 / ST 800
STT 3000
MC Toolkit (MCT404) User Manual
(STT 3000)
LRL (r): Lower Range Limit
URL (r): Upper Range Limit
LRV (r/w): Lower Range Value
URV (r/w): Upper Range Value
PV Units (r/w)
Selection of scaling value (default: inches of H
SV Units (r/w)
Selection of scaling value (C/F)
Sensor Type (r)
Sensor Type associated with the transmitter
Damping (r/w)
Selection of level of digital noise reduction
Span (r)
Process Range (URL - LRL)
Conformity (r/w) (ST 3000 / ST 800)
Selection of conformity to input form: Linear or Square
Root
Linearization (r/w) (STT 3000)
Selection of conformity to input form: Linear or Non
Linear
For more information, refer
to:
Table 14, Table 15, Table
16, Table 17.
PV Type (r/w)
Select: Dual Range
(STDC) or Single Range
or Single Range w/SV
Communication Mode
(r/w) Select: Analog or DE
4 Byte or DE 6 Byte
FS Direction (r)
(Upscale or Downscale;
selection is jumpered in
the transmitter).
Line Filter (r)
Select: 50hz or 60hz.
T/C Fault Detect (r/w):
Select: Enabled or
Disabled.
O@39F)