Aux Input and Output Commands

AUXI? i
AUXO(?)i{,x}
The AUXI? command queries the Aux Input values. The parameter i (1 or 2)
selects an Aux Input and is required. The Aux Input voltages are returned in units
of Volts. The resolution is 1/3 mV. This is a query only command.
The AUXO command sets or queries the Aux Output voltage. The parameter i
(1 or 2) selects an Aux Output and is required. The parameter x is the output
voltage (real number of Volts) and is limited to –10.500
Aux Input and Output 4-17
10.500 .
x
SR844 RF Lock-In Amplifier