The Basic Lock-In - Stanford Research Systems SR844 User Manual

Rf lock-in amplifier
Table of Contents

Advertisement

The Basic Lock-In

This measurement is designed to use the internal oscillator to explore some of the basic lock-in
operations. You should already be familiar with the fundamentals of lock-in detection. See Chapter 2 for a
discussion of the basics of lock-in measurements.
Specifically, you will measure the amplitude of the reference oscillator at various frequencies,
sensitivities, time constants and phases. You will need a BNC cable for this section.
1
Disconnect all cables from the SR844.
If the power is off, turn it on. Wait for self-
tests to complete.
Press Shift then Recall (PRESET) to restore
2
factory presets.
3
Connect REF OUT on the front panel to the
SIGNAL IN with the BNC cable.

The Basic Lock-In

Turn on the unit.
We will start with the unit in its factory preset
configuration.
The factory preset configuration is Internal
Reference mode (shown by the INTERNAL LED)
at 1.00 MHz, shown on the Reference display.
The time constant is 100 ms (shown by the time
constant indicators 1, × × × × 100 and ms above the Time
Constant Up/Down keys) and the sensitivity is 1 V
rms (the indicators are below the Sensitivity
Up/Down keys).
The SR844 reference output (1.0 Vpp nominal
square wave into 50
) is within the unit's
measurement range (1 Vrms) so we can connect it
directly to the input. The SR844 input impedance is
(shown by the 50
set to 50
appropriate for the REF OUT.
The BNC cable has a small phase shift at 1 MHz
(the free-space wavelength is 300 m), so the input
signal should be mostly X (in-phase) with a small Y
(quadrature) component.
The CH1 Display is set to X and should show
something close to 0.54 V. The CH2 display is set
to Y and should show less than
SR844 RF Lock-In Amplifier
1-5
Ω Ω Ω Ω
indicator) which is
±
0.05 V.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents