Save and Recall

Nine setups of the SR844 may be saved in non-volatile memory (setup buffers
1-9). The stored setups include all front panel instrument settings, as well as the
remote interface configurations. The Scan parameters used for manual scans are
saved. The Rel mode data (frequencies, configurations and stored Rel values)
are not stored. The stored setups do not include any signal history or overload
conditions, nor do they include the internal data buffers.
Save
Pressing Save once displays SAVE n where n is the last used setup buffer. A new
setup buffer (1–9) may be selected using the knob. The Reference display shows YES if
buffer n is in use and NO if it is empty.
A second press of the Save key will save the current instrument setup in the chosen
setup buffer. A confirmation message SAVE n DONE is displayed briefly.
Any other keypress will abort the save process and display the message SAVE NOT
DONE.
Recall
Pressing Recall once displays RCAL n where n is the last used setup buffer. Another
setup buffer (1–9) may be selected using the knob. The Reference display shows YES if
buffer n is in use and NO is it is empty.
A second press of the Recall key will recall the instrument setup from the chosen setup
buffer. A confirmation message RCAL n DONE is displayed if the recall operation is
successful. RCAL DATA ERR is displayed if no setup was previously saved in the
selected buffer.
Any other keypress will abort the recall process and display the message RCAL NOT
DONE.
If a remote command is received before the second Recall keypress, the remote
command is processed normally and the instrument continues to wait for either knob
input or the second Recall keypress.
Important!
The recall operation will clear the internal data buffers (see Chapter 4, Data
Storage).
Recall does not affect the stored Rel mode information (see Scan and Rel later in this
chapter). Thus an instrument configuration may be saved at one frequency and
recalled at another frequency, without affecting stored Rel values.
Interface setup parameters are not changed.
PRESET
This key sequence restores the instrument to its factory defaults (see earlier in this
[Shift–Recall]
chapter).
Save and Recall
SR844 RF Lock-In Amplifier
3-31