4. Frequency Accuracy

This test measures the frequency accuracy of the SR844. This tests the accuracy of the frequency counter
inside the unit. The counter is used only in External Reference mode.
Setup
We will use the RF synthesizer (Marconi 2023 or equivalent) to provide the external reference signal.
Connect the synthesizer output to REF IN.
Set the RF synthesizer to:
Frequency:
Amplitude:
Modulation:
Procedure
1) Press Shift then Recall to restore the factory preset instrument settings.
2) Press the keys in the following sequence:
Source
Switch to EXTERNAL Reference mode.
Shift then Freq (Precise Freq)
Display the Precise Frequency on the CH2 and Reference displays.
3) For each frequency in the table below, repeat steps 3.1 through 3.3.
Test Frequencies
1.9993 MHz
1.9997 MHz
3.1) Set the RF synthesizer to the frequency in the table.
3.2) Wait a few until the SR844 UNLOCK indicator is off.
3.3) Record the frequency reading shown on the CH1 and Reference displays. The most
significant 3 digits are on the CH1 display. The least significant 3 digits are on the
Reference display. There are a total of 6 digits in the frequency reading.
4) This completes the Frequency Accuracy test. Enter the results of this test in the test record at the
end of this chapter.
1.9993 MHz
0.0 dBm
OFF
Performance Tests 5-11
SR844 RF Lock-In Amplifier