3-22 CH2 Display and Output
Key features and parameters for the various displayed quantities are shown below.
Quantity
[Unit]
Y [Volts]
θ [Deg]
Ynoise
[Volts]
Ynoise
[dBm]
AUX IN 2
[Volts]
Note 1
Note 2
Note 3
Note 4
Note 5
Offset
User entered offsets can be added to X and Y. These offsets are added before taking
ratios, output time-constant filtering, and computing R and
SR844 RF Lock-In Amplifier
Display
Bar
Range
Graph
Range
±
±
110% f.s.
f.s.
±
±
180°
180°
±
±
110% f.s.
f.s.
±
±
220 dBm
200 dBm
±
±
10 V
10 V
If ratio mode has been selected, the reciprocal of the appropriate input
(1.0V/AUX IN 1 or 1.0V/AUX IN 2) is computed, and both X and Y are
multiplied by this quantity.
the ratio. Ynoise will be scaled in the same proportion as Y.
If ratio mode has been selected, the reciprocal of the appropriate input
(1.0VAUX IN 1 or 1.0V/AUX IN 2) is computed, and both X and Y are
multiplied by this quantity.
unchanged by the ratio. The phase will change by 180° if the ratio input is
negative.
The detection phase may be modified by adjusting the Reference Phase; this
can be used to adjust or null the displayed phase of the input signal, but it
also modifies X and Y (and their offsets). There is no separate phase offset
adjustment.
This shows the worst-case output sample rate for the CH2 analog output.
The sample rate is fastest at the high end of an octave band. See IF
frequency display in the Reference Section for more details.
The X output updates 4 times slower when 18 and 24 dB/oct filtering is
used.
The digital display is always updated at a 2 Hz rate. The bar-graphs are
updated at 64 Hz. Display quantities may be read out via remote interface
or stored in the internal data buffers (see Chapter 4, Data Storage) at a
maximum rate of 512 Hz.
The noise is computed at 512 Hz for all time constants ≤ 30 ms. For longer
time constants the noise is updated 25.6 times per time constant.
Ratio
Offset
Expand
±
Yes
110% f.s. Yes
Note 2
Note 3
Yes
Note 1
No
Yes
Note 1
No
No
No
No
No
θ
is computed from Y/X and is unchanged by
θ
is computed from Y/X and, except for sign, is
θ
.
Max
Sample
Period
µ
22
s, Note 4
µ
88
s, Note 4
1.953 ms,
Note 5
1.953 ms,
Note 5
µ
88
s, Note 4