4 – 14
OFLT(?) z
Example:
OFSL(?) z
Example:

4.4.8 Output commands

OMOD(?) z
Example:
OFSE(?) z
Example:
Output filter time constant
Set (query) the output filter time constant setting to z . Allowable
values are:
z Value
Full-scale sensitivity
500 s
TCMIN 0
TC1MS 1
1 ms
3 ms
TC3MS 2
10 ms
TC10MS 3
30 ms
TC30MS 4
TC100MS 5
100 ms
TC300MS 6
300 ms
1 s
TC1S 7
TC3S 8
3 s
10 s
TC10S 9
30 s
TC30S 10
100 s
TC100S 11
300 s
TC300S 12
OFLT 7
Output filter slope
Set (query) the output filter slope (rollo ) to z (SLOPE6DB 0,
SLOPE12DB 1) .
OFSL SLOPE12DB
Output mode
Set (query) the output mode to z (LOCKIN 0, ACVOLT 1) .
OMOD?
LOCKIN
Output o set enable
Set (query) the output o set mode to z (OFF 0, ON 1) .
OFSE 1