10. Adjustments
Additionally to those ecom-EN2 functions described previously, vari-
ous adjustments can be made in the instrument.
From the main menu, select the sub-menu "Adjustments" and con-
firm with <Enter>.
A selection of modifiable parameters,
adjustable according to the application,
is displayed.
Place the cursor on the desired line and
press <Enter> to call up or modify the
adjustment. The modifiable parameters
are:
Unit (adjustment with cursor keys):
- Calculation of gas concentrations in:
- ppm
3
- mg/m
- mg/kWh (undiluted) = mass concentration per power unit
- mg/MJ (undiluted)
- ppm (undiluted)
3
- mg/m
(undiluted)
Undiluted:
Conversion of the gas concentration on selected reference oxygen:
- mg/kWh and mg/MJ are always calculated on 0% O
- Conversion formula
E
ref
Second unit (adjustment with cursor keys):
-two different unit for one gas concentration possible
O
reference
2
(for ppm und mg/m
- Input of O
2
ecom-EN2
= volume concentration (parts per million)
= mass concentration per volume unit
= mass concentration per power unit
= volume concentration (parts per million)
= mass concentration per volume unit
= E
*
meas
3
- Input after pressing <Enter>):
reference value O
Einheit
Ref. O2
Fuel type
Air pressu
Select
Set clock
Paper feed
Internal
21 – O
2ref
21 – O
2meas
2ref
Adjustments
↵
↑↓
↑↓
↑↓
↑↓
:
basis
2
Page 27