Erasing Data From Memory - AEMC Simple Logger II User Manual

Hide thumbs Also See for Simple Logger II:
Table of Contents

Advertisement

4.6

Erasing Data from Memory

Erasing data from the instrument's memory can only be performed while
in the STANDBY mode.
There are two ways to erase the memory:
Erasing the Memory using the PRESS Button:
1. Press and hold the PRESS button. When the ERASE (RED) LED
lights up, release the button.
This will arm the instrument for an erase operation (when not in record
mode). While armed to erase memory, the RED LED will blink at a fast
rate for a period of five seconds.
2. Press the PRESS button for another 0.5 seconds to start the erase
operation. During the erase operation the LED will blink once every
second. Erasure takes about 20 seconds to complete.
If the button is not pressed within five seconds of arming, the erase oper-
ation will automatically disarm and memory will be maintained. For this
reason, if you do not intend to erase memory, simply wait until the RED
LED stops blinking at the fast rate.
Erasing the Memory using the Data Logger Control Panel:
1. Connect the instrument to the computer, then open the Data Logger
Control Panel.
2. Select Erase Memory from the Instrument Menu.
3. A dialog box will appear asking to confirm the erasure of the instru-
ment's memory. Select Yes to confirm or No to cancel the operation.
Erasing the memory will also clear any stored alarms.