MODEL
ELECTRICAL
Channels
Input
Measurement Range:
(thermocouple dependent)
Resolution
Accuracy (J, K, T, N, E)
Accuracy (R, S)
Temperature
Coefficient:
Maximum Differential
Voltage
Sample Rate
Storage Rate
Storage Modes
Recording Length
Memory
Communication
Power Source**
Battery Life
MECHANICAL
Dimensions
Weight (with battery)
Case
Vibration
Shock
Drop
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
SAFETY & ELECTRO-MAGNETIC COMPATIBILITY
Safety Rating
Protection Degree
CE Approved
22
Two miniature thermocouple connectors
°F
J
-346 to +2192
K
-328 to +2501
T
-328 to +752
N
-328 to +2372
E
-238 to 1742
R
32 to 3212
S
32 to 3212
θ < 1000°F or °C 0.1°F or °C
θ ≤ -148°F/-100°C ± (0.2% of Reading + 1.1°F/0.6°C)
-148°F/-100°C < θ ≤ +212°F/+100°C
± (0.15% of Reading + 1.1°F/0.6°C)
θ > +212°F/+100°C ± (0.1% of Reading + 1.1°F/0.6°C)
θ = 32°F/0°C to +212°F/+100°C
± (0.15% of Reading + 1.8°F/1.0°C)
θ > +212°F/+100°C ± (0.1% of Reading + 1.8°F/1.0°C)
± (0.02% of Reading + 0.03°C)/ °C
or ± (0.02% of Reading + 0.03°F) / °F between operating
temperature of -10°C to +18°C and +28°C to +50°C
Maximum of 8 samples taken at storage interval
Programmable from 5 seconds to 1 per day
Start/Stop, FIFO and Extended Recording Mode* (XRM
15 minutes to 8 weeks, programmable using DataView
240,000 measurement (512KB). Recorded data is stored in non-volatile
memory and retained even if battery is low or removed.
2x1.5V AA (LR6) alkaline batteries
100 hours to > 45 days (dependent on storage rate/recording length)
4.94 x 2.75 x 1.28" (125 x 70 x 32mm)
Up to 85% at 95°F (35°C), Non-condensing
EN61010-1; 50V CAT III; Pollution Degree 2
L642
Two
J
K
T
N
E
R
S
θ ≥ 1000°F or °C 1°F or °C
1V (between inputs)
USB 2.0 optically isolated
7 oz (200g)
UL94-V0
IEC 68-2-6 (1.5mm, 10 to 55Hz)
IEC 68-2-27 (30G)
IEC 68-2-32 (1m)
14° to 122°F (-10° to 50°C)
-4° to 140°F (-20° to 60°C)
2000m
IP40
Yes
°C
-210 to +1200
-200 to +1372
-200 to +400
-200 to +1300
-150 to +950
0 to 1767
0 to 1767
), Alarm
™
®