There are two thresholds for the battery voltage:
The first is used to indicate low battery. The low battery indicator (Blue
LED single blink) will blink when the battery voltage drops below 2.2V.
The second is used to determine when to terminate recording and
turn the unit off. The shutdown threshold is when the battery voltage
drops below 1.7V.
Communication from the instrument to DataView
4.3

Recording Data

The Simple Logger
performed (see the DataView
Once a configuration is written to the instrument, the logger will no longer
need to be connected to DataView
When data is stored in the memory, the user may download the informa-
tion onto a hard-disk.
A scheduled recording will still start even if the logger is in sleep mode.
4.3.1

Starting a Recording Session

A new recording cannot be started if the memory is full.
1. Connect the instrument to the measurement source.
2. Make sure the logger is in STANDBY mode.
3. Press and hold the PRESS button. When the START (GREEN) LED
lights up, release the button.
4. The GREEN LED double-blinks every 5 seconds when the logger is
recording.
It may take a few seconds before the instrument starts recording. The
recording status LED will single blink during this period.