Appendix A ________________________________________________ Values for Internal Monitoring
APPENDIX A

VALUES FOR INTERNAL MONITORING

Status Message Typical
AMBL
Ambient light
-0 V
OFFSET
Measurement
140Hz...
signal offset. The
150Hz
lowest frequency
for measurement
signal
REC. BACKSCATTER
Measurement
signal from the
receiver
contamination
control circuit.
'CHANGE' value
is the instant
signal difference
VAISALA _______________________________________________________________________ 111
The Typical values come from factory testing and are allowed to
change within the given Min/Max limits. The Fault limits give hints
for troubleshooting.
Table 27
Values from STA message for Internal Monitoring
Min. / Max.
Fault
-9 V / +3 V
> 3 V
'RECEIVER
ERROR'
< -9 V
'RECEIVER
SATURATED'
80 Hz /
< 80 Hz
170 Hz
>170 Hz
'OFFSET
ERROR'
Rec.backscatter
change (Instant
backscatter
minus clean
backscatter) >
alarm limit
'BACKSCATTER
HIGH'
Description
Action
Preamplifier is
Change
not working.
PWC22.
The sun is
Check the
shining directly or
orientation of
from reflecting
PWD22.
surface (snow
etc.) to the
receiver.
Drifting may be
1) Check
caused by other
other parts of
disturbing
STA
optical sensors
message.
nearby or other
2) If offset
interference
does not
problems.
remain stable,
change the
control unit
PWC22.
Rising value can
Clean the
be caused by
outside of
contamination of
optical
receiver optics,
surfaces and
snow inside the
remove
receiver hood,
possible
spider net in front
disturbances
of the hood etc.
from optical
Check the
path.