Biotek Cytation 5 Instructions For Use Manual page 91

Cell imaging multi-mode reader
Table of Contents

Advertisement

Parameter
Filters:
Optics Position:
Gain:
Read Speed:
Read Height:
Read Step 3
Detection Method:
Read Type:
Step Label:
Read Wells:
Filter Set:
Filters:
Optics Position:
Gain:
Read Speed:
Read Height:
Read Step 4
Detection Method:
Read Type:
Step Label:
Read Wells:
Filter Set:
Filters:
Optics Position:
Gain:
Cytation 5
Default Setting
EX 485/20 nm, EM 528/20 nm
Top 510 nm
Auto, Use first filter set gain from FIRST Read Step
Normal
Delay after plate movement: 350 msec
Measurements per data point: 50
Lamp Energy: Low
Dynamic Range: Standard
7.00 mm
Fluorescence
Endpoint
"Corners Read"
A1..A3, A10..A12, H1..H3, H10..H12
1 (Green)
EX 485/20 nm, EM 528/20 nm
Top 510 nm
Auto, Scale to High Wells, A3, 30000
Normal
Delay after plate movement: 350 msec
Measurements per data point: 50
Lamp Energy: Low
Dynamic Range: Standard
7.00 mm
Fluorescence
Endpoint
"Linearity Read"
C1..F5
1 (Green)
EX 485/20 nm, EM 528/20 nm
Top 510 nm
Auto, Scale to High Wells, C1, 30000
Fluorescence Liquid Tests |
79