Biotek Cytation 5 Instructions For Use Manual page 118

Cell imaging multi-mode reader
Table of Contents

Advertisement

106
| Specifications
Accuracy, Linearity, Repeatability
Linearity (by liquid dilution)
96-well plate, normal read speed
0–2 OD: +/-1% +/-0.010 OD, delay after plate movement = 100 ms
2–2.5 OD: +/-3% +/-0.010 OD, delay after plate movement = 100 ms
384-well plate, normal read speed
0–2 OD: +/-2% +/-0.010 OD, delay after plate movement = 100 ms
2–2.5 OD: +/-5% +/-0.010 OD, delay after plate movement = 100 ms
96-well and 384-well plate, sweep read speed
0–1 OD: +/-1% +/-0.010 OD
Repeatability (tested with certified neutral density glass/measured by one
standard deviation: 8 measurements per data point)
96-well and 384-well plate, normal read speed
0–2 OD: +/-1% +/-0.005 OD, delay after plate movement = 100 ms
2–2.5 OD: +/-3% +/-0.005 OD, delay after plate movement = 100 ms
96-well and 384-well plate, sweep read speed
0–1 OD: +/-2% +/-0.010 OD
Take3 Plate
Detection Limit, 260 nm dsDNA
< 5 ng/µL
BioTek Instruments, Inc.

Advertisement

Table of Contents
loading

Related Products for Biotek Cytation 5

Table of Contents