Clean The Objectives - Biotek Cytation 5 Instructions For Use Manual

Cell imaging multi-mode reader
Table of Contents

Advertisement

Do not touch the filters with your bare fingers!
1. Open the access door on the front of the instrument. Slide the filter cube out
of its compartment.
2. Inspect the glass filters for speckled surfaces or a "halo" effect. This may
indicate deterioration due to moisture exposure over a long period of time.
If you have any concerns about the quality of the filters, contact your
BioTek representative.
3. Using cotton balls or lens-cleaning tissue moistened with a small amount of
high-quality alcohol, clean each filter by lightly stroking its surface in one
direction.
4. Use a magnifying glass to inspect the surface; remove any loose threads left
from the cotton ball.
5. Replace the filter cube and close the door.

Clean the Objectives

Applies only to models with the imaging module.
The objectives used in the Cytation 5 should be cleaned when necessary using
optical-grade swabs or lens paper moistened with lens cleaning solution or deionized
water.
Do not rub the lens.
Materials
• Air puffer
• Tweezers
• Magnifying glass
• Lens cleaning tissue
• Optical-grade swabs
• Cleaning solvent
Recommended Cleaning Solvents
HyperClean (hexamethyldisoloxan and ethanol),
available from Olympus
Isopropyl alcohol, 70%/30% with deionized water
Methyl alcohol, 70%/30% with deionized water
Ethyl alcohol, 70%/30% with deionized water
Cytation 5
Clean the Objectives |
Non-recommended Cleaning
Solvents
Methyl ethyl ketone (MEK)
Dimethyl ketone (acetone)
49