46
| Maintenance
5. Wipe all exposed surfaces of the dispense module (if used).
6. Wipe all exposed surfaces of the gas controller module (if used).
7. If detergent was used, wipe all surfaces with a cloth moistened with water.
8. Use a clean, dry cloth to dry all wet surfaces.
Models with a dispenser: If the Tip Priming Trough overflows or other
spills occur inside the instrument, wipe the carrier and the surface
beneath the carrier with a dry cotton cloth. The internal chamber and
probes are not customer-accessible. If overflow is significant, contact
BioTek's Technical Assistance Center with any questions about your
particular model.

Inspect/Clean Mirrors

Applies only to Cytation 5 models with filter-based fluorescence capabilities.
We recommend inspecting/cleaning the mirrors and polarizing filters (if equipped)
annually, especially if the filter cube has been opened or changed.
These optical elements are delicate and should be handled as carefully as possible.
The glass and anti-reflective (AR) coated surfaces will be damaged by any contact,
especially by abrasive particles.
surface.
However, if performance indicators or obvious defects in the mirrors or
filters suggest cleaning them, here are some guidelines:
• Use of oil-free dry air or nitrogen under moderate pressure is the best method
for removing excessive debris from an optical surface. If the contamination is
not dislodged by the flow of gas, please follow the cleaning instructions below.
In most cases, it is best to leave minor debris on the
BioTek Instruments, Inc.