3.5 PI MEASUREMENT

MΩ-1000V
MΩ-1000V
Before the
measurement
(2 Presses)
During the
measurement
(4 Presses)
After the measurement
(6 Presses)
The values of PI and DAR are calculated as follows:
PI = R 10 min / R 1 min (2 values to be recorded during a measurement
lasting 10 min).*
DAR = R 1 min / R 30 s (2 values to be recorded during a measurement
lasting 1 min).
*For the calculation of the PI, the times of 1 and 10 min can be modified by the
user, if required, for a particular application. See § 3.5.1.
16
Set the switch to one of the insulation
measurement positions.
Start the measurement with a long press (> 2 s)
on the START/STOP button. The long press will
be acknowledged by an audible beep.
The measurement starts the default duration
of 10 min. A countdown will display the
time remaining. The measurement will stop
automatically.
Press the DISPLAY button to display:
Voltage present on the device to be tested.
Test voltage.
Leakage current present.
Measurement time remaining.
Instantaneous insulation resistance value.
Test voltage.
Current flowing through the resistance being
measured.
Value of PI (available at the end of 10 min).
Value of DAR (available at the end of one min).
Test voltage generated during the measurement.
PI and DAR.
Duration of the measurement.
Insulation resistance value just before the
measurement was stopped.
Current that flowed in the resistance measured.
Voltage present on the device being tested.
Capacitance.
Surface leakage current.
Megohmmeter Model 6505 - User Manual