Service and maintenance
6.1
Maintenance of analysis module
6.1.1
Removal of analysis module
Procedure
Explosion hazard
If the gas chromatograph is mounted in a hazardous area, observe the explosion
Before opening the housing in the hazardous area, switch off the power supply and then
Following the waiting time:
1. Loosen the 1.5-mm hexagon socket-head bolt
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
WARNING
protection regulations.
always wait for at least 65 minutes.
– When the surface temperatures inside the gas chromatograph have cooled down to
a temperature corresponding to temperature class T4.
– When the electric charge on the internal components has dropped to a voltage
permissible for explosion protection.
turns.
③
②
in the top cover
6
by two complete
77