Siemens MicroSAM Operating Instructions Manual page 110

Table of Contents

Advertisement

Technical data
Sample and injection
Sample streams
Calibration sample streams
Phase
Permissible sample pressure
Sample flow
Max. sample temperature
Particulates
Required filtration
Material wetted by the sample
Injection
Control
Injection volume adjustable using switching times
Detectors, calibration and performance data
Detector type
Cell volume
Calibration
Permissible performance of carrier gases
Repeatability for measuring ranges 2 ... 100%
Repeatability for measuring ranges 0.2 ... 1%
Repeatability for measuring ranges < 0.1%
Detection limit for measuring ranges > 1000 ppm
Smallest measuring range
Linear range
Cycle time
Ambient temperature influence
Influence of vibrations
Mean time to repair (MTTR)/
Mean time between failures (MTBF)
108
3 for devices of release version 4 onwards
1 for devices up to release version 3
Note:
The release version is present on the rating plate at "AS".
1
Gaseous
10 ... 60 kPa above atmospheric pressure
20 to 100 ml/min
80 °C
< 1 μm
Degree of separation 99.99 % for 1 μm particles
Stainless steel, fused silica, polyamide
Valveless live injection, i.e. no switching valves required in oven
area.
with multifunctional membrane valve
From 2 to 50 μl
TCD, max. 8 sensors
0.02 μl
Manual or automatic, single-level or
multi-level (only MicroSAM)
± 0.5% of full-scale value (application-dependent)
± 1% of full-scale value (application-dependent)
± 2% of full-scale value (application-dependent)
1% of full-scale value (application-dependent)
1000 ppm (application-dependent)
Typically ≥ 10
4
Typically 30 ... 240 s
Negligible
Negligible
< 1 hour/3 years (without consumables)
Operating Instructions, 06/2012, C79000-G5376-C560-07