3. Adjust the M14 nut flush as shown in the following photo at
4. Screw the M14 nut by hand onto the guide pin. Make sure that the adjustment is retained.
5. Tighten the M10 lock nut using the ring spanner A/F 10.
6. Rotate the M14 nut clockwise using the A/F 21 ring spanner until the 2.0 mm thread
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
CAUTION
Damage to the connection
When installing the analysis module, the packing glands of the module are sealed by the
pneumatic connector. Only use the supplied ring spanner. If you apply too much force
by using a different ring spanner with a longer lever arm, the connection will be
damaged.
becomes visible.
Service and maintenance
6.1 Maintenance of analysis module
.