3.6.4
Thermal release in the analysis module
A thermal release is fitted in the analysis module. This switches off the power supply for the
complete device as soon as the maximum permissible temperature of the analysis module of
155 °C is exceeded.
Note
The thermal release used cannot be reset or replaced. If the thermal release is triggered,
you must replace the analysis module.
3.6.5
Measuring sequence
The measuring sequence for live and eLive injection is always as follows:
1. Preparation of measurement (Page 35)
2. Purge sample loop (Page 39)
3. Injecting (Page 41)
4. Empty sample loop completely (Page 43)
5. Purge sample loop with sample (Page 44)
6. Cut - start (Page 45)
7. Cut - end (Page 46)
8. Backflushing (Page 47)
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
Description
3.6 Functions
33
Need help?
Do you have a question about the MicroSAM and is the answer not in the manual?