Setup
Press or to scroll through the run time presets on the LED.
Run Time Mode
Timed Run (defaults: 15, 10, 5, 2, and 1 min)
Intermitt ent Sampling
Continuous Run
The user may change run time presets. See Advanced Setup, Programming Your Own
Run Time Presets.
Timed Run
Run time accumulates only while the pump is running.
Timed Run mode is designed to perform one sample for a
predetermined time from 1 to 999 minutes (selected from
presets). The user starts the pump running and the pump
automatically stops running after the sample time has
elapsed.
1. Press or to scroll to the desired run time preset.
2. Press [
] to select the desired preset. The pump will
start running and the LED will display a count down to
zero (run time remaining below one minute is displayed
in seconds with a colon). The pump will stop and beep
four times. The display will fl ash Done.
3. Press
to return to the run time preset display. If the
user wishes to repeat the sample, press [
User Options During Timed Run Sampling
Viewing batt ery status
Press
. Display will automatically return to run time
remaining after fi ve seconds or press
Hold mode
Press [
pump and timer in Hold. Run time remaining and Hold
will display alternately. Press [
continue the sample run.
Terminating a sample and resett ing the pump
Press
while in Hold. Sample Reset (rSEt) will fl ash on
the LED.
To terminate the sample and reset the pump: Press [
To continue the sample run: Press
[
].
] while the sample is running to place the
LED Display
t xx
Intt
Cont
].
again.
] while in Hold to
to return the display to Hold. Press
15
Digital display:
Timed Run 2 minutes
:
Digital display:
Run time remaining is
2 seconds
Digital display:
Sample run completed
].
Need help?
Do you have a question about the QuickTake 30 and is the answer not in the manual?