• To navigate presets and displays:
 or 
Press
• To select a run time preset and run the pump:
Scroll to desired run time preset.
Select preset by pressing [
Selecting the
Available Presets
Press  or 
• To run the pump:

Press [
].
• To place the pump in hold:

Press [
].
• To repeat a sample run:
From Done, reset (rSEt), or sampling error (SErr), press [
time preset. Press [
• To turn off LED:
The pump automatically turns off LED after it is idle for 4 minutes. User can
reactivate the LED by pressing and holding
• To interrupt a run, terminate a sample, or reset a pump:
From Hold, press
pump or
to ignore the reset and return to Hold.
• To set fl ow or calibrate:
See Calibration and Sett ing Flow for details.
1. Connect a calibrator to the pump inlet.
2. Scroll to any run time preset. Press and hold CAL for 2 seconds to enter
calibration mode (pump will start running).
3. Press  or  to change fl ow rate until desired fl ow appears on the
calibrator (fl ow rate does not display on LED).
4. Press
and Stor will display. Press [
changes.
to scroll through presets or displays.

]. The pump will start to run.
Press 

] to sample.
. Sample reset (rSEt) will display. Press [
.

] to save sett ing or
11
Press 
5
4
3
2
1
CONT
INTT
END
999
998
997

] to return to run

] to reset the
to ignore