Programming and Calibration
301IRFS Specifications
301D2 Sensor Specifications
Sensing technology:
Sensor power requirements:
Length of lines:
Operating Temperature Range:
Operating Humidity Range:
Operating Altitude:
Enclosure:
50
Electrochemical (toxic)
Catalytic combustion (combustible)
Diffusion fuel cell (oxygen)
Toxic: 7 - 9Vdc, 25mA@ 8 Vdc
Combustible: 7-9 Vdc, 100mA@ 8 Vdc
Up to 160m (500 feet ) between
301EM and sensor (toxic and
combustible)
Toxic: -40 to 40°C (-40°F to 100°F)
Combustible: -40 to 50°C (-40°F to
112°F)
0% to 95% RH, non-condensing
Up to 3000m (9843 feet )
Class I, Division 1, Groups B, C, D
301EM User Manual