To Clean the CI Source

Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Solvents
• Ultrasonic bath
Preparation
1
2
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
3
5975 Series MSD Operation Manual for MassHunter
Large (8650-0030)
Small (8650-0029)
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Disassemble the ion source. See
page 167.
Collect the following parts to be cleaned for a CI source:
Repeller
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
Clean the parts as described in
"To Disassemble the CI Source"
"To Clean the EI Source"
6
CI Maintenance
on
on page 146.