Agilent Technologies 5975 Operation Manual page 93

Series msd
Hide thumbs Also See for 5975:
Table of Contents

Advertisement

FB = Feedthrough Board
Repeller
(red wire
from FB)
Filament 1
(white wires
from FB)
Ion focus lens
(orange wire
from FB)
Entrance lens
(blue wire
from FB)
Figure 18
Ion source wiring
2 Check the side plate O-ring.
5975 Series MSD Operation Manual for MassHunter
Make sure the O-ring has a very light coat of Apiezon L high vacuum grease.
If the O-ring is very dry, it may not seal well. If the O-ring looks shiny, it has
too much grease on it. (Refer to the 5975 Series MSD Troubleshooting and
Maintenance Manual for lubricating instructions.)
Operating in Electron Impact (EI) Mode
3
Ion source
heater wires
Ion source
sensor wires
Filament 2
(black wires
from FB)
93

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents