3
4
The flow control module
The CI reagent gas flow control module
flow of reagent gas into the CI GC/MSD interface. The flow module consists of
a mass flow controller (MFC), gas select valves, CI calibration valve, shutoff
valve, control electronics, and plumbing.
The back panel provides Swagelok inlet fittings for methane (CH4) and one
OTHER reagent gas. The software refers to them as Gas A and Gas B,
respectively. If you are not using a second reagent gas, cap the OTHER fitting to
prevent accidental admission of air to the analyzer. Supply reagent gases at
25 to 30 psi (170 to 205 kPa).
5975 Series MSD Operation Manual for MassHunter
The system evacuates the gas lines for 6 minutes, then turns on the
selected gas (A or B). This is to reduce cross-mixing of the gases in the
lines.
Enter the reagent gas flow setpoint in the Flow field. This value is entered
as a percentage of maximum flow rate. The recommended flow is 20% for a
PCI source and 40% for an NCI source.
The flow control hardware remembers the flow setting for each gas. When
either gas is selected, the control board automatically sets the same flow
that was used for that gas the last time.
To begin reagent gas flow, select Shutoff Valve.
The system turns off the present gas flow while leaving the shutoff valve
(Figure 21
on page 112) open. This is to remove any residual gas in the
lines. Typical evacuation time is 6 minutes and then the shutoff valve is
closed.
Operating in Chemical Ionization (CI) Mode
(Figure 21
and
Table
17) regulates the
4
111