2 Remove the septum cap from the column OUTLET side and uncoil 2 column
Figure 3
3 Attach three column clips (part number G2630-20890) to the column cage
5975 Series MSD Operation Manual for MassHunter
loops. See
Figure
3.
Column with 2 uncoiled loops
as follows:
Attach one clip onto the back of the 1 o'clock cross-member piece of the
column cage.
Attach two clips onto the front of the 3 o'clock cross-member piece of the
column cage.
These clips will help provide appropriate orientation of column ends for
their insertion into the GC inlet and MSD interface.
Installing GC Columns
1 o'clock cross-member
3 o'clock cross-member
2
35