4 Trace the wires for the ion source heater and temperature sensor to the
5 Remove the thumbscrews that hold the ion source in place.
6 Pull the ion source out of the source radiator.
The analyzer operates at high temperatures. Do not touch any part until you are sure
WARNING
it is cool.
Source feedthrough board
Ion source
Thumbscrews
Source heater and temperature sensor wires
Source radiator
Figure 26
Removing the ion source
5975 Series MSD Operation Manual for MassHunter
feedthrough board. Disconnect them there.
General Maintenance
5
143