4
Operating in Chemical Ionization (CI) Mode
The shutoff valve prevents contamination of the flow control module by
atmosphere while the MSD is vented or by PFTBA during EI operation. The
MSD monitors will reflect On as 1 and Off as 0 (see
Gas A
(methane)
supply
Gas B
(other)
supply
Figure 21
Reagent gas flow control module schematic
Table 17
Flow control module state diagram
Result
Gas A flow
Gas A
Open
Gas B
Closed
MFC
On
setpoint
112
Gas A
select valve
Gas B
select valve
Gas B flow
Purge
with Gas A
Closed
Open
Open
Closed
On
setpoint
On
Mass
flow
controller
Calibration
valve
Restrictor
Calibration
vial
Purge
with Gas B
Closed
Open
100%
On
100%
5975 Series MSD Operation Manual for MassHunter
Table
17).
CI ion
source
Shutoff
valve
GC/MSD
interface
GC column
Pump out
Standby,
flow module
vented, or EI
mode
Closed
Closed
Closed
Closed
On
100%
Off
0%