Agilent Technologies 5975 Operation Manual page 147

Series msd
Hide thumbs Also See for 5975:
Table of Contents

Advertisement

If insulators are dirty, clean them with a cotton swab dampened with reagent-grade
CA UT IO N
methanol. If that does not clean the insulators, replace them. Do not abrasively or
ultrasonically clean the insulators.
Procedure
The filaments, source heater assembly, and insulators cannot be cleaned
CA UT IO N
ultrasonically. Replace these components if major contamination occurs.
1
2
3
4
5
All of these solvents are hazardous. Work in a fume hood and take all appropriate
WARNING
precautions.
5975 Series MSD Operation Manual for MassHunter
If the contamination is serious, such as an oil backflow into the analyzer,
seriously consider replacing the contaminated parts.
Abrasively clean the surfaces that contact the sample or ion beam.
Use an abrasive slurry of alumina powder and reagent-grade methanol on
a cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also
abrasively clean the discolorations where electrons from the filaments
enter the source body.
Rinse away all abrasive residue with reagent-grade methanol.
all
Make sure
abrasive residue is rinsed
methanol becomes cloudy or contains visible particles, rinse again three
times.
Separate the parts that were abrasively cleaned from the parts that were
not abrasively cleaned.
Ultrasonically clean the parts (each group separately) for 15 minutes. For
dirty parts, use all three solvents in the order shown, cleaning 15 minutes
with each of the following solvents:
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Methanol (reagent-grade)
For routine cleaning, cleaning with methanol is sufficient.
General Maintenance
before
ultrasonic cleaning. If the
5
147

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents