Installing GC Columns
Figure 10
7 Loosen the interface nut and push the column an additional 3–5 cm into
8 Make a clean cut of the column so that now only 3–5 cm protrudes into the
9 Clean the outside of the free end of the column with a lint-free cloth
10 Adjust the column so that it protrudes 1 to 2 mm into the analyzer chamber
48
22–28 cm from 3 o'clock clip to GC/MSD interface nut
Oven door opened and closed
the analyzer chamber.
analyzer chamber.
moistened with methanol.
past the end of the GC/MSD interface, and hand tighten the nut. See
Figure
11.
Make sure the position of the column does not change as you retighten the
nut.