Test Button Operation - AEMC Megohmmeters 6522 User Manual

Input impedance 400komh
Hide thumbs Also See for Megohmmeters 6522:
Table of Contents

Advertisement

> 1s
TEST
The
voltage (>70V).
6. At the end of the measurement, release the TEST button. The
instrument stops generating the test voltage and discharges the device
being tested. The
system under test has fallen below 70V.
NOTE: Do not disconnect the leads and do not start any measure-
ment while the symbol
When you release the TEST button, the measurement results remain displayed
until the next measurement, or the HOLD button is pressed, or the instrument is
turned OFF.

2.2.2.1 TEST Button Operation

Pressing the TEST button starts an insulation measurement. In normal mode, the
test voltage is generated for as long as the button is pressed. When the button is
released, the measurement stops.
In
mode, press the test button once to start the measurement, then press it a
second time to stop; there is no need to keep the button pressed. However, if you
do not stop the measurement, it will stop automatically after 15 minutes.
In timed test mode (
measurement. The test will stop automatically at the end of the defined test
duration time.
symbol indicates that the instrument is generating a hazardous
symbol is displayed until the voltage on the
, DAR, or PI) press the TEST button once to start the
G
is displayed.
M
V
M

Advertisement

Table of Contents
loading

This manual is also suitable for:

Megohmmeters 6524

Table of Contents