Storage Memory (Model 6524); Operating Environment; Voltage Measurement - AEMC Megohmmeters 6522 User Manual

Input impedance 400komh
Hide thumbs Also See for Megohmmeters 6522:
Table of Contents

Advertisement

3.2.7 Storage Memory (Model 6524)

Maximum number of recordings stored in memory: 300

3.3 Operating Environment

3.3.1 Voltage Measurement

Influencing
Parameter
Temperature
Relative
humidity
Frequency
Supply
voltage
Common
mode
rejection in
AC 50/60 Hz
Megohmmeter Model 6522/6524
Range of
influence
20 to + 55°C
(-4 to 131°F)
20 to 80%
RH
15.3 to
800Hz
6.6 to 9.6V
0 to 600V
AC
Quantity
influenced
V, F
V, F
V
V, F
V
Influence
Typical
Maximum
0.3% R/10°C
(0.3% R/18°F
1% R + 2 ct
1%
2% R + 1 ct
0.1% R + 2 ct
50dB
+ 1 ct
+ 1 ct)
40dB
33