In each measurement mode, the third threshold can be replaced by a user-
defined value. To do this:
1. Press the ► button while the threshold value is displayed.
2. The > symbol starts blinking; you can change it to < by pressing the
button. This symbol indicates the direction of the alarm threshold: < for a
low threshold and > for a high threshold.
3. To change the threshold setting, press the ► button to navigate to the
first digit, and then use the button to change its value.
4. Use the ► and buttons to select and change the other digits in the
threshold value, as well as the units of measurement.
5. When finished setting the threshold, press the ► button to validate the
setting.
2.1.2.3 Viewing Alarms
When the measurement is below a low alarm threshold or above a high alarm
threshold, the instrument emits a continuous audible signal and the LCD
indicates the threshold crossed:
When checking continuity, this enables you to determine whether or not the
continuity measurement is less than 2Ω simply by listening, without looking at the
display. You can similarly check insulation quality.
<
ALARM
>
ALARM
M
k
M