Frequency Measurement; Insulation Measurement - AEMC Megohmmeters 6522 User Manual

Input impedance 400komh
Hide thumbs Also See for Megohmmeters 6522:
Table of Contents

Advertisement

3.2.2 Frequency Measurement

Measurement Range
Resolution
Intrinsic uncertainty

3.2.3 Insulation Measurement

Specific reference condition: Capacitance in parallel on resistance = null
Measurement Ranges per Model
Test Voltage
50V
100V
250V
500V
1000V
Accuracy
Test Voltage
(V
)
T
Measurement
Range
Resolution
Accuracy
For all test voltages, when the insulation resistance is ≤ 2GΩ the intrinsic
uncertainty is ± (3% R + 2 ct).
Bar Graph
Measurement Range
Resolution
Intrinsic uncertainty
*When the measurement range is exceeded, the whole bar graph is displayed.
Model 6522
50kΩ to 10GΩ
100kΩ to 20GΩ
200kΩ to 40GΩ
50V - 100V - 250V - 500V - 1000V
10 to 999kΩ
4.00 to
and 1.000 to
39.99MΩ
3.999 MΩ
1kΩ
10kΩ
V
V
T
V
T
V
T
V
= 1000V: ± (3% R + 2 ct + 0.1%/GΩ)
T
15.3 to 399.9V
0.1Hz
± (1% R + 2 ct)
40.0 to
399.9MΩ
100kΩ
= 50V: ± (3% R + 2 ct + 2%/GΩ)
T
= 100V: ± (3% R + 2 ct + 1%/GΩ)
= 250V: ± (3% R + 2 ct + 0.4%/GΩ)
= 500V: ± (3% R + 2 ct + 0.2%/GΩ)
0.1MΩ - 200GΩ*
9 segments per decade
± (5% R + 1 segment)
400 to 700V
1Hz
± (1.5% R + 1 ct)
Model 6524
10kΩ to 10GΩ
20kΩ to 20GΩ
50kΩ to 50GΩ
100kΩ to 100GΩ
200kΩ to 200GΩ
400 to
4.00 to
3999MΩ
39.99GΩ
1MΩ
10MΩ
40.0 to
200.0GΩ
100MΩ

Advertisement

Table of Contents
loading

This manual is also suitable for:

Megohmmeters 6524

Table of Contents