Setting An Alarm Threshold - AEMC Megohmmeters 6522 User Manual

Input impedance 400komh
Hide thumbs Also See for Megohmmeters 6522:
Table of Contents

Advertisement

Model 6524:
To turn OFF the alarm buzzer while it is sounding, press the HOLD button. To
deactivate an active alarm function, press the

2.1.2.2 Setting an Alarm Threshold

While
is displayed indicating the alarm function is active, you can change
the alarm threshold by pressing the  button (except during insulation
measurements). For each testing mode, there are three pre-defined threshold
values:
■ Continuity: < 2Ω, <1Ω and <0.5Ω
■ Resistance: >50kΩ, >100kΩ and >200kΩ
■ Insulation:
50V : <50kΩ, <100kΩ and <20kΩ
100V : <100kΩ, <200kΩ and <400kΩ
250V : <250kΩ, <500kΩ and <1MΩ
500V : <500kΩ, <1MΩ and <2MΩ
1000V : <1MΩ, <2MΩ and <4MΩ
Megohmmeter Model 6522/6524
The alarm is available in insulation, resistance, and continuity
measurement modes. Pressing the
alarm. The
symbol is displayed, along with the threshold
value.
<