RF Analyzer
Tests and Adjustments
N9912A
Performance Tests (Keysight N7841A Software Package)
Performance Tests (Keysight N7841A Software Package)
The performance tests section of the Keysight N7841A software package verifies the electrical performance
of your N9912A analyzer. The software automatically configures your analyzer to execute the performance
tests.
The N7841A software package is not included with the analyzer; it must be ordered separately. It can be
downloaded from the Internet and a license purchased on-line, or it can be ordered on a CD-ROM. Go to
www.keysight.com/find/calibrationsoftware
for further information.
There are:
• Four tests for the general functionality of the analyzer.
• Seven additional tests for the SA (spectrum analyzer) functions.
Total test time will vary from 7 to 36 minutes depending upon options installed.
Details for these tests and recommended troubleshooting information can be found in the N7841A help file.
General Tests
Total test time for these tests is 8 minutes.
• Frequency Accuracy Test
• Error Terms Test (CAT Functional)—This test includes an adjustment (see
"Adjustments (Keysight
N7841A Software Package)" on page
3-8) and should be performed every 12 months.
• High/Low Output Power Output Test (CAT Functional)
• Crosstalk S21 Test (CAT Functional)
Spectrum Analyzer Specific Tests
Total test time for these tests is 29 minutes.
• Absolute Amplitude Accuracy Test
• IF Frequency Response Test (Functional)
• Frequency Response with Preamp Off Test
• Frequency Response with Preamp On Test
• Displayed Average Noise Level with Preamp On Test
• Noise Sidebands Test
Service Guide N9912-90003
3-7