Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG
5.3
P10 normal view for children with significant dose reduction, P10
A artifact-reduced, P10 C with a constant magnification factor of
1.25
PAN
PAN
CEPH
1260
P10
10
11,5s
Ready for exposure
59 87 594 D 3352
D 3352.201.01.18.02
Plus
DS/Ceph
TS
64kV
8mA
?
Active sensor area CCD PAN
6.5 x 127.0 mm,
for upper or lower jaw:
6.5 x 65.0 mm
5.3 P10 normal view for children with significant dose reduction, P10 A
To preselect the program, use the arrow keys (3): + (to
count forward) or – (to count backward), depending on
6
the starting position; for program sequence see section
3.2.
8
To preselect the exposure type, touch the program dis-
play (6). A continuous loop will then display all programs
one after another, P10, P10 A, P10 C, P10...
3
In the "Half-view and individual quadrant selection" sub-
menu at the top of column (8) you can preselect whether
you want to generate a right or left half-view, or a maxil-
lary or mandibular view (or an individual quadrant view
in the full version) (see section 3.2 General touchscreen
functions).
P 10 normal view (status)
preferably for children
•
Yellow bite block or contact segment
or
chin rest with bite block rod and bite block or bar.
•
Head inclination using the FH.
The line displayed on the head symbol on the touch-
screen indicates the reflection line of the light local-
izer.
•
Optional presetting for malocclusion
•
Optional presetting for jaw shape
•
Automatic selection of the slice width in accor-
dance with temple support settings for different den-
tal arches. Radiation time depends on set temple
support width.
P10 A normal view (status) artifact-reduced
preferably for children
•
Settings as for P10
•
To prevent metal artifacts in the condylar and molar
region and to reduce shadowing caused by the op-
posite jaw.
P10 C normal view (status) with a constant
magnification factor of 1.25
preferably for children
e.g. for implantology
•
Settings as for P10
5 Program group panoramic images
33