Setting Calibration

The
indicates calibration mode or calibration setup .
Before beginning a calibration, prepare the electrode or probe according to the
instructions received with the electrode or probe . The calibration standards, buffers,
or calibration sleeve should also be prepared prior to the calibration . After preparing
the electrode or probe per manufacturer's instructions, connect the electrode or
probe to the meter . Basic information on proper calibration for each measurement
technique is found in the appropriate section of this manual .
General Navigation for Selecting Calibration
1 . In measurement mode press
pointed toward the measurement mode to the calibrated:
pH, ORP , ISE • Conductivity • DO
2 . Press
/
be calibrated .
3 . Then press
Thermo Scientific Orion Star
until the appropriate icon is lit for the measurement mode to
A
– pH/ORP/ISE
B
– Conductivity Calibration
C
– DO Calibration
to begin the selected calibration .
Series Meter
until the arrow icon on the left of the screen is
pH icon – pH Calibration
RmV icon – ORP Calibration
ISE icon or no icon – ISE Calibration
mS/cm, μS/cm, mg/L, ppt or MΩ-cm icon
% sat or mg/L icon
No icon – Barometer Calibration
Menu Setup
EN-23