Menu Setup
Continuous, Timed, or AUTO-READ
The following table describes in greater detail the abbreviations shown on the
display screen .
Description
Continuous, Timed,
or AUTO-READ
Measurement Selection
Timed Reading Setting
To change settings, simply use
Pressing
will confirm the change of a value and end the flashing display .
Orion Star Series meters allow you to select your preferred timing for measurement
intervals .
EN-20
Display
rEAd
tYPE
COnt
rEAd
tInE
00:00
and
In continuous mode the instrument is constantly making
measurements and updating the screen . To print and log
the measurement in this mode press
In timed mode the instrument is constantly making
measurements, updating the display, and will automatically
log and print data at a pre-selected time interval .
In the setup menu enter the minutes and seconds
between timed readings by pressing
change the value of the flashing digit . Press
change the digit to be scrolled .
The minimum time interval is 5 seconds . Acceptable
time ranges are:
Measurement Selection
Range of alues
COnt, AUtO, tImE
Timed, continuous, or
AUTO-READ; automatically
prints and logs in AUTO-READ
and timed modes
00:05 - 99:61
Timed readings in minutes
and seconds
/
to move between options .
Thermo Scientific Orion Star
Method
Default
Specific
alue
Yes
AUtO
Yes
01:00
.
/
to
to
Series Meter