Calibration:store; Calibration:test - HP E1418A User Manual

8/16-channel d/a converter module
Hide thumbs Also See for E1418A:
Table of Contents

Advertisement

CALibration:STORe

Parameters
Comments

CALibration:TEST?

Parameters
Comments
Chapter 3
loads the stored non-volatile memory calibration constants
CALibration:STORe
with the calibration constants currently in use by the module.
None.
This command sets the calibration constants in non-volatile memory if the CAL
Store Enable Jumper is properly set (see page 159). Non-volatile memory has
finite life, based upon the number of writes. Use this command sparingly.
At power-on, calibration constants stored in non-volatile memory are loaded into
the module where they are used during output operations.
When this command is received, the checksum is verified using the constants
currently in use by the module. If the checksum does not verify, the non-volatile
memory is not updated and an error is generated.
Refer to Appendix D, "Voltage/Current Output Adjustment", starting on page 157,
for additional information about this command.
Related Commands:
performs an internal rough check of the calibration of each
CALibration:TEST?
channel and returns an integer indicating the channel calibration condition.
Calibration is checked to approximately ± 2%.
Returns: unsigned 16-bit integer (uint16).
Following the completion of this query all outputs are disabled and set to 0 Volts
or 0 Amps based on the channel mode.
This query returns a 16-bit integer indicating the calibration status of each channel.
Bits set to 0 indicate no calibration errors were found.
Bits set to 1 indicate a channel with questionable calibration.
Bits 0–15 correspond to channels 1–16, respectively.
If the trigger system is waiting for a trigger when this command is received, an
error is generated.
CALibration:RESet
HP E1418A SCPI Command Reference
CALibration
68