CALibration

CALibration:CONFigure:RESistance?

Parameters
Comments

CALibration:RESet

Parameters
Comments
67
HP E1418A SCPI Command Reference
CALibration:CONFigure:RESistance?
calibration of the internal calibration resistor and returns a
readiness of the module.
Returns: signed 16-bit integer (int16).
The internal calibration resistor should be calibrated before calibrating the
channels in current mode.
This query configures the module for a 4-wire resistance measurement at the CAL
output terminals when the calibration point is set to
DIAGnostic:CALibration:OUTPut[:MODE] CALBus
This query may also be used to set an external calibration resistor value when
calibrating at each channel output.
This query is used together with the
calibrate the current calibration resistor.
Refer to Appendix D, "Voltage/Current Output Adjustment", starting on page 157,
for additional information about this command.
Related Commands:
CALibration:RESet
reloads, for module use, the calibration constants stored in
non-volatile memory.
None.
This command, with respect to the calibration constants, performs the same action
as power-on.
At power-on, or following this command, calibration constants stored in
non-volatile memory are loaded into the module where they are used during
output operations.
If the trigger system is waiting for a trigger when this command is received,
an error is generated.
This command opens all output relays.
Related Commands:
configures the module to perform
1
= ready,
CALibration:VALue:RESistance
CALibration:VALue:RESistance
CALibration:STORe
0
or
1
indicating the
0
= not ready.
(following a
CALB
*RST
command).
command to
Chapter 3
or