Sourcen:current[:Level][:Immediate][:Amplitude] - HP E1418A User Manual

8/16-channel d/a converter module
Hide thumbs Also See for E1418A:
Table of Contents

Advertisement

SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]
Parameters
Comments
Chapter 3
n
SOURce
:CURRent[:LEVel][:IMMediate][:AMPLitude] <
immediately outputs the current set in value at channel n when the output relay is closed.
Parameter
Parameter
Name
n
keyword
SOURce
value
numeric (float 32) or
<
>
discrete
In the keyword,
SOURce
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The
value
parameter is a floating point 32-bit number in the range of -0.02 A
<
>
to +0.02 A.
sets -0.02 Amps.
MIN
sets +0.02 Amps.
MAX
sets 0 Amps.
DEF
This command immediately programs the specified current on channel n. The
output state of the channel must be enabled for the current to appear at the output
terminals.
If the channel is in the voltage mode when this command is received, an error is
generated.
Related Commands:
SOURce
n
:FUNCtion:MODE CURRent
Type
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
n, the valid range of n is:
SOURce
n
:CURRent[:LEVel][:IMMediate][:AMPLitude]?
,
OUTPut
HP E1418A SCPI Command Reference
SOURce
value
| MIN | MAX | DEF>
Range of
Values
1 to 8 or 1 to 16
n
[:STATe]
Default
1
0
,
84