SOURce
n
:VOLTage[:LEVel][:IMMediate][:AMPLitude]
Parameters
Comments
Chapter 3
n
SOURce
:VOLTage[:LEVel][:IMMediate][:AMPLitude] <
immediately outputs the voltage set in value at channel n when the output relay is closed.
Parameter
Parameter
Name
n
keyword
SOURce
value
numeric (float 32) or
<
>
discrete
In the keyword,
SOURce
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The
value
parameter is a floating point 32-bit number in the range of -16 Volts
<
>
to +16 Volts.
sets -16 Volts.
MIN
sets +16 Volts.
MAX
sets 0 Volts.
DEF
This command immediately programs the specified voltage on channel n. The
output relay for that channel must be closed for the voltage to appear at the output
terminals.
This command generates an error if channel n is in the current output mode.
Related Commands:
SOURce
n
:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Type
-16 V to +16 V or
MIN | MAX | DEF
n, the valid range of n is:
n
SOURce
:FUNCtion:MODE VOLTage
HP E1418A SCPI Command Reference
SOURce
value
| MIN | MAX | DEF>
Range of
Values
1 to 8 or 1 to 16
,
,
OUTPut
n
[:STATe]
Default
1
0