SOURce
n
:CURRent[:LEVel]:TRIGgered[:AMPLitude]
Parameters
Comments
Chapter 3
n
SOURce
:CURRent[:LEVel]:TRIGgered[:AMPLitude] <
outputs the current set in value at channel n when the trigger conditions are satisfied and
the output relay is closed. Use this command to synchronize multiple channel outputs.
Parameter
Parameter
Name
keyword
SOURce
n
numeric (float 32) or
<
value
>
discrete
In the keyword,
SOURce
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The
<
value
>
parameter is a floating point 32-bit number in the range of -0.02 A
to +0.02 A.
sets -0.02 Amps.
MIN
sets +0.02 Amps.
MAX
DEF
sets 0 Amps.
This command sets channel n to output the current specified following a valid
trigger condition. Following this command, the trigger system must be enabled
using an
INITiate
command. See the
for additional information.
If the module is in a wait-for-trigger state (
command is received, an error is generated.
The output state of the channel must be enabled and the trigger event satisfied for
the current to appear at the output terminals.
If the channel is in the voltage mode when this command is received, an error is
generated.
Related Commands:
SOURce
n
:FUNCtion:MODE CURRent
Type
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
n, the valid range of n is:
TRIGger
subsystem discussion on page 54
INIT
n
SOURce
:CURRent[:LEVel]:TRIGgered[:AMPLitude]?
,
OUTPut
HP E1418A SCPI Command Reference
SOURce
value
| MIN | MAX | DEF>
Range of
Values
1 to 8 or 1 to 16
has been received) when this
n
[:STATe]
Default
1
0
,