Code
Error Messages
+3008 Invalid Calibration State
+3009 Invalid Calibration Channel
+3010 Calibration aborted
+3011 Checksum error on Calibration
Memory
+3012 Channel cannot be programmed to
this mode, check jumpers
+3013 Waiting for Trigger
+3014 Illegal while initiated
+3015 Non-volatile memory write failed
Appendix C
Potential Cause(s)
An invalid sequence of calibration commands was received.
For example, a CAL
n
:CONF:VOLT command followed by a
CAL
n
:VAL:CURR command.
An invalid channel number,
CAL
n
:CONF or CAL
n
:VAL commands.
A command, other than one of the CALibration commands, was
received during the calibration process.
Calibration constants in RAM are corrupted. Recalibration is
needed.
An attempt to set the channel output mode when the P/J jumper
is set to Jumper selectable and the mode is set by the V/I
jumper.
An attempt to change a channel output value or mode was
made after the trigger system in the wait-for-trigger state.
An attempt to program a channel was made when the channel
is in the wait-for-trigger state.
Indicates a hardware failure in non-volatile memory.
HP E1418A Error Messages
n
, was requested in the