CALibration

CALibration:VALue:RESistance

Parameters
Comments
69
HP E1418A SCPI Command Reference
CALibration:VALue:RESistance <
as the internal calibration resistor constant by the module.
Parameter
Parameter
Name
numeric (float 32)
value
<
>
The new calibration constant is used by the module but not stored in non-volatile
memory.
This command is used together with the
query to calibrate the internal current calibration resistor.
The internal calibration resistor should be calibrated before calibrating channels in
current mode. When calibrating at the CAL connections and using the internal
calibration resistor, the value need only be entered once to calibrate all channels.
If using an external resistor to calibrate current mode, you must calibrate at each
channel output. External resistances in the range of 50 Ω to 500 Ω are recommended.
Refer to Appendix D, "Voltage/Current Output Adjustment", starting on page 157,
for additional information about this command.
Related Commands:
value
enters a new measured value to be used
>
Type
Must be > 0 ohms
CALibration:CONFigure:RESistance?
CALibration:STORe
,
CALibration:CONFigure:RESistance?
Range of
Values
Chapter 3
Default