SOURce
n
:VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Parameters
Comments
Chapter 3
n
SOURce
:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
outputs the voltage set in value at channel n when the trigger conditions are satisfied
and the channel output relay is closed. Use this command to synchronize multiple
channel outputs.
Parameter
Parameter
Name
keyword
SOURce
n
numeric (float 32) or
<
value
>
discrete
•
In the keyword,
SOURce
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
The
value
parameter is a floating point 32-bit number in the range of -16 Volts
<
>
to +16 Volts.
sets -16 Volts.
MIN
MAX
sets +16 Volts.
sets 0 Volts.
DEF
•
This command sets channel n to output the voltage specified following a valid
trigger condition. Following this command, the trigger system must be enabled
using an
command. See the
INITiate
additional information.
•
The channel output relay must be closed for the voltage to appear at the output
terminals.
•
If the module is in a wait-for-trigger state (
command is received, an error is generated.
•
If the channel is in the current mode when this command is received, an error is
generated.
•
Related Commands:
SOURce
n
:FUNCtion:MODE VOLTage
Type
-16 V to +16 V or
MIN | MAX | DEF
n, the valid range of n is:
subsystem discussion, page 54, for
TRIGger
INIT
SOURce
n
:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
HP E1418A SCPI Command Reference
SOURce
value
| MIN | MAX | DEF>
Range of
Values
1 to 8 or 1 to 16
has been received) when this
Default
1
0
,